Strontium in a Dicronite® Coating? Why EDX and XRF Can Give a False Positive
A customer's lab reported strontium on a part coated with Dicronite® DL-5®. Here is why X-ray analysis of a 0.5 µm film can name an element that is not there.
Interactive graphic
What the Instrument Actually Measures
Switch the analysis method or the spectrum view, or select a numbered point or a card to see the details.
Analysis method
EDX spectrum, 1.1 to 2.1 keV
Tap a number or swipe to see the next point.
The question
Strontium Detected in the Coating?
A customer ran analyses on parts coated with Dicronite® DL-5® and detected strontium in the coating. They asked whether strontium is expected in the coating, or whether it pointed to contamination.
The answer is neither. The strontium reading is a false positive, an artifact of the limits of X-ray analysis on a coating this thin. This article explains why, based on Lubrication Sciences International's Technical Bulletin 121525.
The analysis was most likely carried out with an industry-standard X-ray technique such as EDX or XRF. These methods are widely used and can be extremely valuable, but this case shows the fundamental limits they face on micron and sub-micron coatings deposited on a metal substrate.
Reason 1
X-Ray Methods See Through a 0.5 µm Coating
Dicronite® is deposited at about 0.5 µm thick. At that thickness, conventional X-ray techniques cannot analyze the coating on its own. The X-rays pass through the coating into the substrate below, so the result is a combined, proportional analysis of the coating and the base material, not the coating chemistry alone.
SEM-EDS / EDX
Energy-dispersive X-ray spectroscopy. Readily detects tungsten and sulfur in the coating, but reaches well past it.
~1–3 µm deepXRF
X-ray fluorescence. Reaches much deeper into the part than EDX.
~5–50 µm deepXPS
X-ray photoelectron spectroscopy. Surface-sensitive, so it can characterize the coating without interference from the substrate.
Nanometers deepAES
Auger electron spectroscopy. Also surface-sensitive, probing only the top nanometers of the part.
Nanometers deep
Any element detected this way on a 0.5 µm Dicronite® coating is strongly influenced, if not dominated, by the substrate and any contamination near the surface. Strontium found this way cannot automatically be attributed to the coating.
Reason 2
Three Peaks Closer Than the Detector Can Separate
In this case the substrate contained no strontium either. The rest of the answer lies in how close together the X-ray energies of tungsten, silicon and strontium are.
- 1.739 keVSi KαSilicon, from the substrate
- 1.775 keVW MαTungsten, from the coating
- 1.806 keVSr LαStrontium, not present in the part
What happens during EDX analysis
- Tungsten (W) from the coating is detected, as expected.
- Silicon (Si) from the substrate is detected at the same time, because the interaction volume extends through the coating into the substrate.
- The W and Si signals partly overlap in the low-energy region, and the analysis software tries to separate them.
- Because of software limits, the combined signal is assigned to strontium, the nearest element in this closely spaced group, even though no strontium is present.
The result is a best-fit false positive, a known artifact of standard EDX systems.
An analogy
Green, Blue and a Color That Is Not There
Think of the tungsten signal as a green peak and the silicon signal as a blue peak. Because the two peaks are so close together, the detector sees a blend of green and blue. The software, trying to name that blend, matches it to the closest color in its library, the one that stands for strontium. The result is a false positive, even though no strontium is present in the substrate or the coating.
The right test
Analyzing a Thin Film Accurately
Accurate elemental analysis of a coating about 0.5 µm thick needs a surface-sensitive technique, such as X-ray photoelectron spectroscopy (XPS) or Auger electron spectroscopy (AES). These probe depths on the nanometer scale and can characterize the coating without interference from the substrate.
Bulk techniques such as EDS/EDX and XRF are inherently influenced by the substrate, which can lead to false positives, exactly as in this case.
Quality assurance
Why We Are Confident Dicronite® Contains No Strontium
- Dicronite® raw powder undergoes chemical analysis every quarter
- The coating is tested, approved and certified to the latest AMS 2530 aerospace specification for tungsten disulfide films
- The strontium reading is explained by the limits of the X-ray analysis equipment
The outcome
The Customer's Response
The customer thanked the team for confirming there was no strontium in the coating. They had suspected as much from their SEM results and wanted confirmation, and found that the explanation matched the results they had seen.
Summary
A False Positive, Not Contamination
On a 0.5 µm Dicronite® coating, EDX and XRF mostly measure the substrate. Silicon from the substrate and tungsten from the coating produce peaks only 30–40 eV apart, closer than a standard EDX detector can resolve, and the software's best fit can name strontium even though none is present.
With quarterly analysis of the raw powder and certification to AMS 2530, Dicronite® is strontium-free. To analyze the coating itself, use a surface-sensitive method such as XPS or AES.